[PRINT] - Advances to next step if span must be set.
[TARE] - Establishes a new zero calibration setting without changing the span
calibration previously set. The display will show CAL d. This feature is disabled if
the Add ALL message is displayed.
888888 Test Weight
Apply known weight to scale. Fractional or decimal weights are not acceptable
only whole numbers. Use the numeric keyboard to enter the applied weight. Press
the [ENTER] button to advance to the next step. If Add ALL was displayed in the
previous step, the system will use the Full Scale Capacity as the test weight value
and automatically proceed to the next step.
15 SEC Span Setting Fifteen Second Delay
Counts down from 15 to zero to capture the span setting. If the [SELECT] button is
pressed during the time-out period, the meter will immediately terminate the delay
and ignore the motion criterion to capture the weight. Allow at least three (3)
seconds after weight has been applied to the scale before pressing [SELECT] or the
weight captured may be erroneous.
CAL d Calibration Done
Indicates that the calibration procedure is complete.
[PRINT] - Advances to next step.
[7] - Decreases span setting.
[9] - Increases span setting.
FCd File Changed Data
Changes will be saved or discarded.
0 - Save the entered values in memory and use again on power up.
1 - Discard present values and restore previously entered values.
2 - Initialize the EEPROM (U14) to the factory settings.
[PRINT] - Advance to the next step.
Page 37